Z-0424-2019 Class II Terminated

Recalled by AROA Biosurgery, LTD. — Lower Hutt

FDA device recall Z-0424-2019 was initiated by AROA Biosurgery, LTD. on Apr 6, 2018 and is designated Class II. Reason for recall: Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over… The recall status is terminated (terminated Apr 15, 2019). Affected quantity: 881 total.

Recall details

Product Type
Devices
Report Date
Nov 21, 2018
Initiation Date
Apr 6, 2018
Termination Date
Apr 15, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
881 total

Product description

OviTex 1S Reinforced BioScaffold 10x20cm, Part Number F10256-1020G

Reason for recall

Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.

Distribution pattern

The products were distributed to the following US states: AL, CA, FL, IN, MA, MI, NH, and NY.

Code information

ERT-6J08 ERT-7J08 ERT-7K08

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0424-2019
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0424-2019. Report a data issue: hello@regdatalab.com.