Z-0432-2018 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
February 7, 2018
Initiation Date
August 7, 2017
Termination Date
March 29, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
638 units

Product Description

Affixus Hip Fracture Nail, LH 130 DEG 11MM X 280MM

Reason for Recall

During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.

Distribution Pattern

US nationwide distribution. Also Portugal and the Netherlands.

Code Information

Item No. 814611280; Lot No. (Exp. Date) UDI: 459270 (05/18/2027) UDI (01)00887868039775 (17)270518 (10)459270; 858780 (06/16/2027) UDI (01)00887868039775 (17)270616 (10)858780; 972840 (06/15/2027) UDI (01)00887868039775 (17)270615 (10)972840