Z-0446-2021 Class II Ongoing

Recalled by Shanghai United Imaging Healthcare Co., Ltd. — Shanghai

FDA device recall Z-0446-2021 was initiated by Shanghai United Imaging Healthcare Co., Ltd. on Aug 5, 2020 and is designated Class II. Reason for recall: A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive… The recall status is ongoing. Affected quantity: 5.

Recall details

Product Type
Devices
Report Date
Nov 18, 2020
Initiation Date
Aug 5, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5

Product description

Computed Tomography X-Ray System, Model uCT 550, Rx, CE, GTIN: 06971576831036 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

Reason for recall

A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.

Distribution pattern

US Nationwide distribution including in the state of TX.

Code information

Serial Numbers: 353020, 353021, 353023, 353045, 353049

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0446-2021
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Shanghai United Imaging Healthcare Co., Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0446-2021. Report a data issue: hello@regdatalab.com.