Z-0447-2021 Class II Ongoing

Recalled by Shanghai United Imaging Healthcare Co., Ltd. — Shanghai

Recall Details

Product Type
Devices
Report Date
November 18, 2020
Initiation Date
August 5, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5

Product Description

Computed Tomography X-Ray System, Model uCT 760, Rx, CE, GTIN: 06971576831043 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

Reason for Recall

A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.

Distribution Pattern

US Nationwide distribution including in the state of TX.

Code Information

Serial Numbers: 600229, 600230, 600232, 600271, 600282