Z-0458-2018 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

FDA device recall Z-0458-2018 was initiated by Zimmer Biomet, Inc. on Aug 7, 2017 and is designated Class II. Reason for recall: During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed… The recall status is terminated (terminated Mar 29, 2019). Affected quantity: 638 units.

Recall details

Product Type
Devices
Report Date
Feb 7, 2018
Initiation Date
Aug 7, 2017
Termination Date
Mar 29, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
638 units

Product description

Affixus Hip Fracture Nail, RH 125 DEG 9MM X 400MM

Reason for recall

During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.

Distribution pattern

US nationwide distribution. Also Portugal and the Netherlands.

Code information

Item No. 814309400; Lot No. (Exp. Date) UDI: 437960 (05/20/2027) UDI (01)00887868038044 (17)270520 (10)437960; 893070 (06/14/2027) UDI (01)00887868038044 (17)270614 (10)893070

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0458-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Zimmer Biomet, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Zimmer Biomet, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Zimmer Biomet, Inc. →

Source: openFDA record for Z-0458-2018. Report a data issue: hello@regdatalab.com.