Z-0460-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 13, 2023
- Initiation Date
- November 3, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 941 devices
Product Description
QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-6X3, Order Number G22655; used to dilate strictures of the biliary tree
Reason for Recall
Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.
Distribution Pattern
US
Code Information
UDI/DI 00827002226555 (each), 10827002226552 (case), Lot Numbers: W4721265, W4737428, W4740108, W4741851, W4742110, W4742520, W4742521, W4743198, W4744020, W4744440, W4744875, W4749508, W4752283, W4756468, W4720846, W4721267, W4721796, W4722643, W4722984, W4723305, W4723845, W4737430, W4737767, W4738715, W4739351, W4742522, W4742526, W4743195, W4743196, W4744021, W4744023, W4744443, W4744874, W4744876, W4748041, W4748965, W4749745, W4750014, W4750263, W4751108, W4751109, W4751510, W4751905, W4751907, W4753815, W4753817, W4754120, W4754121, W4754253, W4756473, W4758166, W4758167, W4758929, W4758930, W4760296, W4760714, W4761223, W4762154