Z-0466-2022 Class II Terminated

Recalled by Ion Beam Applications S.A. — Louvain-la-neuve, N/A

Recall Details

Product Type
Devices
Report Date
January 19, 2022
Initiation Date
November 29, 2021
Termination Date
February 20, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5 devices

Product Description

Proteus 235; Version: PTS-8 versions before PTS-8.7.2

Reason for Recall

Proton Therapy System (PTS) software can be started in the clinical environment (clinical user) while some processes are still running in the test environment (tcs user). Initiating the startup of the clinical processes in the environment is not prevented even if some processes of the test environment (tcs) are still running. It may lead to situations where the system is using test processes without notifying the clinical user. If the versions of the processes are different between both environments, this could have an impact on patient treatments.

Distribution Pattern

IN USA: VA, OK, NJ, TN OUS: Czech Republic.

Code Information

PAT.108 (US), PAT.110 (US), PAT.113 (US), PAT.114 (EU), SAT.119 (US).