Z-2434-2021 Class II Ongoing

Recalled by Ion Beam Applications S.A. — Louvain-la-neuve, N/A

FDA device recall Z-2434-2021 was initiated by Ion Beam Applications S.A. on Jun 14, 2021 and is designated Class II. Reason for recall: Users may be misled by the popup message displayed by the Proton Therapy System when the Oncology Information System appears disconnected, may cause mistreatment The recall status is ongoing. Affected quantity: 27 worldwide, 11 in U.S.

Recall details

Product Type
Devices
Report Date
Sep 15, 2021
Initiation Date
Jun 14, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
27 worldwide, 11 in U.S.

Product description

Proteus 235- designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation

Reason for recall

Users may be misled by the popup message displayed by the Proton Therapy System when the Oncology Information System appears disconnected, may cause mistreatment

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of FL, OK, VA, IL, NJ, TN, WA, LA, TX, MI and the countries of Germany, Italy, Russia, Sweden, India, Japan, France, England, Spain.

Code information

PAT.006 (US), PAT.107 (EU), PAT.108 (US), PAT.110 (US), PAT.112 (US), PAT.113 (US), PAT.114 (EU), PAT.115 (EU), SAT.116 (US), SAT.118 (RU), SAT.119 (US), SAT.120 (EU), SAT.122 (EU), SAT.123 (US), SAT.125 (IN), SAT.126 (US), SAT.132 (EU), SBF.101 (EU), SBF.103 (JP), SBF.104 (JP), SBF.105 (US), SBF.107 (EU), SBF.109 (UK), SBF.110 (UK), SBF.113 (US), SBF.115 (UK), SBF.117 (EU).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2434-2021
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2434-2021. Report a data issue: hello@regdatalab.com.