Z-1945-2024 Class II Ongoing

Recalled by Ion Beam Applications S.A. — Ottignies-Louvain-La-Neuve, N/A

FDA device recall Z-1945-2024 was initiated by Ion Beam Applications S.A. on May 10, 2024 and is designated Class II. Reason for recall: Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances The recall status is ongoing. Affected quantity: 2 units.

Recall details

Product Type
Devices
Report Date
Jun 12, 2024
Initiation Date
May 10, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product description

IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Reason for recall

Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances

Distribution pattern

Worldwide - US Nationwide distribution in the state of Virginia and the country of India.

Code information

UDI: SAT140: (01)05404013814015(11)190507(21)SAT140 SAT136: (01)05404013801138(11)201105(21)SAT136 Serial Numbers: SAT136, SAT140 PTS-11 versions from PTS-11.1 and PTS-12 versions, with C230 accelerator

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1945-2024
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1945-2024. Report a data issue: hello@regdatalab.com.