Z-1945-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 12, 2024
- Initiation Date
- May 10, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2 units
Product Description
IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
Reason for Recall
Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances
Distribution Pattern
Worldwide - US Nationwide distribution in the state of Virginia and the country of India.
Code Information
UDI: SAT140: (01)05404013814015(11)190507(21)SAT140 SAT136: (01)05404013801138(11)201105(21)SAT136 Serial Numbers: SAT136, SAT140 PTS-11 versions from PTS-11.1 and PTS-12 versions, with C230 accelerator