Z-1205-2022 Class II Ongoing

Recalled by Ion Beam Applications S.A. — Louvain-la-neuve, N/A

Recall Details

Product Type
Devices
Report Date
June 8, 2022
Initiation Date
April 20, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 units

Product Description

Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation

Reason for Recall

First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of IL, VA, WA and the country of Germany.

Code Information

UDI-DI: (01)05404013801138 Serial Numbers: PAT108 (US), PAT112 (US), SAT116 (US). PAT107 (EU).