Z-0468-2015 Class II Terminated

Recalled by Summit Industries Inc. — Chicago, IL

FDA device recall Z-0468-2015 was initiated by Summit Industries Inc. on Oct 28, 2014 and is designated Class II. Reason for recall: The welds on J700 tube stands may be insufficient resulting in tube stand breakage, damage to the X-Ray machine, and/or patient or healthcare provider injury. The recall status is terminated (terminated May 21, 2015). Affected quantity: 32.

Recall details

Product Type
Devices
Report Date
Dec 3, 2014
Initiation Date
Oct 28, 2014
Termination Date
May 21, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
32

Product description

The J700 Floor Mounted Tube Stand (FMTS) is intended to support and position the diagnostic x-ray tube housing assembly for a medical radiographic procedure. The product consists of a vertical column (with counterbalanced vertical carriage, counterweight and pulley system), horizontal floor car and a floor mounted track to ride on.

Reason for recall

The welds on J700 tube stands may be insufficient resulting in tube stand breakage, damage to the X-Ray machine, and/or patient or healthcare provider injury.

Distribution pattern

US Distribution including the states of: AZ, CO, IL, KY, LA, NE, NY, OH, PA, TN, TX, VA, WA and WV.

Code information

Model Numbers: J700-01, J700-02, J700-03, J700-04

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0468-2015
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Summit Industries Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0468-2015. Report a data issue: hello@regdatalab.com.