Z-0470-2013 Class II Terminated
FDA device recall Z-0470-2013 was initiated by Stryker Howmedica Osteonics Corp. on Jul 27, 2012 and is designated Class II. Reason for recall: Offset Bushings associated with the lots identified were mismarked: the rotational reference numbers on the Tibial Offset Bushing were marked counterclockwise instead of clockwise; the Femoral Offset… The recall status is terminated (terminated Sep 30, 2015). Affected quantity: 8.
Recall details
- Product Type
- Devices
- Report Date
- Dec 12, 2012
- Initiation Date
- Jul 27, 2012
- Termination Date
- Sep 30, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 8
Product description
Stryker Orthopaedics TIBIAL OFFSET BUSHING TRIATHLON REVISION INSTRUMENTS REF 6543-2-600;NON-STERILE Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 USA A Subsidiary of Stryker Corp. Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 Meyzieu Cedex France Made in USA The Offset Bushings are parts of the Triathlon TS system. They are used during trialing to assess the appropriate magnitude and orientation of offset between the Implant and the intramedullary Stem. The feedback from the Offset Bushing allows for the selection of the corresponding Offset Adaptor, which allows for optimal bone coverage, compensating for differences between the Implant and bone canal axes.
Reason for recall
Offset Bushings associated with the lots identified were mismarked: the rotational reference numbers on the Tibial Offset Bushing were marked counterclockwise instead of clockwise; the Femoral Offset Bushing did not include rotational markings.
Distribution pattern
Worldwide Distribution.
Code information
Class I, 510(k) exempt Catalog Number: 6543-2-600 Lot Number: P1S43, P2A31