Z-0479-2014 Class II Terminated

Recalled by Aesculap, Inc. — Center Valley, PA

FDA device recall Z-0479-2014 was initiated by Aesculap, Inc. on Sep 10, 2013 and is designated Class II. Reason for recall: Complaints were received for the GN161 Bipolar Foot Control reporting that the bipolar energy did not stop after release of the foot control. The recall status is terminated (terminated Oct 20, 2014). Affected quantity: 85.

Recall details

Product Type
Devices
Report Date
Dec 18, 2013
Initiation Date
Sep 10, 2013
Termination Date
Oct 20, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
85

Product description

Aesculap GNI161 Bipolar Foot Control Product Usage: The Aesculap Bipolar Coagulator is intended for use in surgery to generate electrical power for bipolar instruments. Bipolar coagulators are used in neurosurgery, ENT surgery, urology, laparoscopy, and plastic surgery.

Reason for recall

Complaints were received for the GN161 Bipolar Foot Control reporting that the bipolar energy did not stop after release of the foot control.

Distribution pattern

Worldwide Distribution - USA Nationwide and countries of: Austria, Germany, Japan, and Taiwan.

Code information

all GN161's distributed between 4/28/2009 and 5/22/2013.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0479-2014
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Aesculap, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0479-2014. Report a data issue: hello@regdatalab.com.