Z-0479-2022 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- January 19, 2022
- Initiation Date
- August 27, 2021
- Termination Date
- February 9, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1 unit
Product Description
Software BeamAdjust version 2.2, Catalog No: S070009 Software VeriSoft version 8.0, Catalog No: S080032
Reason for Recall
When a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or too-high measured absolute dose which affects patient plan verification results
Distribution Pattern
US Nationwide distribution in the state of CO.
Code Information
S/N :AU2001094-1 UDI: EPTWS0700090 VeriSoft ;EPTWS0700090 BeamAdjust