Z-0479-2022 Class II Terminated

Recalled by PTW NORTH AMERICA CORPORATION — Boonton, NJ

FDA device recall Z-0479-2022 was initiated by PTW NORTH AMERICA CORPORATION on Aug 27, 2021 and is designated Class II. Reason for recall: When a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or… The recall status is terminated (terminated Feb 9, 2024). Affected quantity: 1 unit.

Recall details

Product Type
Devices
Report Date
Jan 19, 2022
Initiation Date
Aug 27, 2021
Termination Date
Feb 9, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product description

Software BeamAdjust version 2.2, Catalog No: S070009 Software VeriSoft version 8.0, Catalog No: S080032

Reason for recall

When a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or too-high measured absolute dose which affects patient plan verification results

Distribution pattern

US Nationwide distribution in the state of CO.

Code information

S/N :AU2001094-1 UDI: EPTWS0700090 VeriSoft ;EPTWS0700090 BeamAdjust

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0479-2022
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0479-2022. Report a data issue: hello@regdatalab.com.