Z-0479-2022 Class II Terminated
FDA device recall Z-0479-2022 was initiated by PTW NORTH AMERICA CORPORATION on Aug 27, 2021 and is designated Class II. Reason for recall: When a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or… The recall status is terminated (terminated Feb 9, 2024). Affected quantity: 1 unit.
Recall details
- Product Type
- Devices
- Report Date
- Jan 19, 2022
- Initiation Date
- Aug 27, 2021
- Termination Date
- Feb 9, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1 unit
Product description
Software BeamAdjust version 2.2, Catalog No: S070009 Software VeriSoft version 8.0, Catalog No: S080032
Reason for recall
When a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or too-high measured absolute dose which affects patient plan verification results
Distribution pattern
US Nationwide distribution in the state of CO.
Code information
S/N :AU2001094-1 UDI: EPTWS0700090 VeriSoft ;EPTWS0700090 BeamAdjust
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0479-2022
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab