Z-0481-2024 Class II Ongoing
FDA device recall Z-0481-2024 was initiated by Heartware, Inc. on Oct 31, 2023 and is designated Class II. Reason for recall: Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm may sound and… The recall status is ongoing. Affected quantity: 126 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 13, 2023
- Initiation Date
- Oct 31, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 126 units
Product description
PUMP 1104 HVAD (HeartWare Ventricular Assist Device) IMPLANT KIT EN JA, Model Number 1104JP
Reason for recall
Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm may sound and the recommended troubleshooting actions and (2) instructions relating to the useful life of the HVAD system components
Distribution pattern
Worldwide
Code information
GTIN none, Serial Numbers: HW25172, HW25169FSCA, HW25170FSCA, HW25174FSCA; GTIN 00763000170967, Serial numbers: HW42722, HW42771, HW42285, HW41696, HW41675, HW40953, HW41848, HW41847, HW41316, HW41828, HW41824, HW41803, HW41670, HW41271, HW41701, HW41674, HW41791, HW41792, HW41804, HW41252, HW41256, HW41700, HW41699, HW41783, HW41317, HW41309, HW41673, HW41254, HW40651, HW40665, HW40663, HW41672, HW41308, HW41269, HW41544, HW41270, HW41214, HW41314, HW40417, HW41263, HW41264, HW41265, HW41047, HW41260, HW41266, HW41251, HW41259, HW41258, HW40579, HW41048, HW41262, HW41045, HW40961, HW40958, HW40956, HW41050, HW41255, HW40881, HW40789, HW40957, HW40960, HW40951, HW40659, HW40602, HW40816, HW40954, HW40950, HW40662, HW40664, HW40580, HW40668, HW40613, HW40652, HW40666, HW40667, HW40660, HW40658, HW40612, HW40420, HW40657, HW40607, HW40552, HW40609, HW40599, HW40424, HW40423, HW40426, HW40600, HW40465, HW40553, HW40554, HW40383, HW40603, HW40653, HW40655, HW40656, HW40473, HW40555, HW40472, HW40422, HW40480, HW40564, HW40565, HW40413, HW40479, HW40557, HW40415, HW40476, HW40478, HW40428, HW40429, HW40477, HW40419, HW40421, HW40425, HW40431, HW40471, HW40474, HW40475, HW40412, HW40414, HW40418
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0481-2024
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
Need the full FDA record for Heartware, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.