Z-0482-2021 Class II Terminated

Recalled by Viewray, Inc. — Mountain View, CA

Recall Details

Product Type
Devices
Report Date
December 2, 2020
Initiation Date
October 20, 2020
Termination Date
October 13, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
35 units

Product Description

ViewRay MRIdian Linac System: Model No. 20000, CE, Rated Supply Voltage - 380/480 VAC, 3 - Frequency Range (Hertz) - 50/60 Hz, Rated Input in Amperes - 210 A, UDI. The MRIdian system and the MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.

Reason for Recall

There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user.

Distribution Pattern

Worldwide Distribution: US (nationwide) to states of: CA, NY, FL, IL, MI, MO, OR, WI and OUS countries of: France, Germany, Italy, United Kingdom, Taiwan, Netherlands, Israel, and Denmark.

Code Information

All Lots/Serial Numbers