Z-0491-2015 Class II Terminated
FDA device recall Z-0491-2015 was initiated by Integra LifeSciences Corp. on Sep 29, 2014 and is designated Class II. Reason for recall: Some specific lots of Selector Tubing may potentially leak irrigation fluid during use if not placed on the Selector handpiece in a specific manner. The recall status is terminated (terminated May 27, 2015). Affected quantity: 6,358 tubing kits.
Recall details
- Product Type
- Devices
- Report Date
- Dec 10, 2014
- Initiation Date
- Sep 29, 2014
- Termination Date
- May 27, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6,358 tubing kits
Product description
Integra Selector Kit Sterile single use device A bi-lumen, sterile single use accessory used for the Selector Integra Ultrasonic Surgical Aspirator system. One of the tubing lumen transmits irrigation fluid from the console to the handpiece of the system, and the other tubing lumen transmits aspirated fluid and tissue back to the system console. The aspirator is intended to be used to facilitate the removal of cellular and other unwanted soft tissue. The system provides selective tissue disintegration with simultaneous irrigation and aspiration. The system has been designed for use by surgeons in the areas of neurosurgery and gastroenterology including laparoscopic procedures. Its use in other procedures is regarded as experimental and is thus subjecto to appropriate local regulatory approval
Reason for recall
Some specific lots of Selector Tubing may potentially leak irrigation fluid during use if not placed on the Selector handpiece in a specific manner.
Distribution pattern
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Australia, Bulgaria, Canada, Germany, Great Britain, Greece, Hungary, Indonesia, Israel, Italy, Japan, Mexico, Portugal, Spain, Sweden, Thailand, Turkey, and Wales.
Code information
Lot Numbers: 130849;1132912;1132913;1132914;1133101;1133102;1133103;1133104;1133105;1133495;1134005;1134006;1134007;1134120;1134121;1134122;1134124;1134126;1134128;1134130;1134131;1134132;1134133;1134620;1134621;1134669;1134670;1134671;1134671;1134694;1134695;1140357;1141561;1141562;1141645;1141754;1141563;1141564;1141647;1142352;1142353;1142409; & 1142410
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0491-2015
- openFDA file date
- Nov 16, 2026