Z-0493-2019 Class II Terminated

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

Recall Details

Product Type
Devices
Report Date
November 28, 2018
Initiation Date
August 30, 2018
Termination Date
October 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Total: 29,982,150 BD cups (US); 97,800,900 BD cups (OUS)

Product Description

KIT URIN CUP PLC 16X100 10.0 UA YEL, Cat. No. 364981: Urine Collection Kit for collection and transport of urine specimens for urinalysis

Reason for Recall

BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.

Distribution Pattern

Illinois

Code Information

All lots within expiry