Z-0501-2025 Class II Ongoing

Recalled by OrganOx Ltd — Oxford, N/A

FDA device recall Z-0501-2025 was initiated by OrganOx Ltd on Oct 15, 2024 and is designated Class II. Reason for recall: Battery movement led to damage of one of the battery cells causing it to short to the battery retention bracket The recall status is ongoing. Affected quantity: 20 units.

Recall details

Product Type
Devices
Report Date
Nov 27, 2024
Initiation Date
Oct 15, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20 units

Product description

OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Number: D0003 The OrganOx metra is a transportable device intended to be used to sustain donor livers destined for transplantation in a functioning state for a total preservation time of up to 12 hours.

Reason for recall

Battery movement led to damage of one of the battery cells causing it to short to the battery retention bracket

Distribution pattern

California, Texas, Ohio, Florida, New York, Massachusetts, Virginia, Oklahoma, DC, South Carolina and Georgia.

Code information

UDI-DI: 05060462240029 Serial numbers 511, 512, 514, 515, 516, 517, 518, 519, 520, 522, 523, 524, 525, 527, 528, 529, 530, 531, 533, 535, 536, 537

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0501-2025
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for OrganOx Ltd? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for OrganOx Ltd: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about OrganOx Ltd →

Source: openFDA record for Z-0501-2025. Report a data issue: hello@regdatalab.com.