Z-0511-2020 Class II Terminated

Recalled by Neocis Inc. — Miami, FL

FDA device recall Z-0511-2020 was initiated by Neocis Inc. on Oct 4, 2019 and is designated Class II. Reason for recall: The Fiducial Array may mismatch their measurement files resulting in failed landmark check. The recall status is terminated (terminated Oct 5, 2020). Affected quantity: 3 units.

Recall details

Product Type
Devices
Report Date
Nov 27, 2019
Initiation Date
Oct 4, 2019
Termination Date
Oct 5, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 units

Product description

Neocis Guidance System Fiducial Array - Product Usage: The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intraoperative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. Fiducial Arrays are plastic parts that contain metal beads, which serve as the fiducial CBCT markers.

Reason for recall

The Fiducial Array may mismatch their measurement files resulting in failed landmark check.

Distribution pattern

US Nationwide distribution in the states of AR, OK, NY.

Code information

Serial Numbers: FA-0128-02, FA-0129-02, FA0130-02

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0511-2020
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0511-2020. Report a data issue: hello@regdatalab.com.