Z-0516-2022 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-0516-2022 was initiated by Siemens Medical Solutions USA, Inc on Oct 12, 2021 and is designated Class II. Reason for recall: Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system… The recall status is terminated (terminated Apr 11, 2024). Affected quantity: 72 units.

Recall details

Product Type
Devices
Report Date
Feb 2, 2022
Initiation Date
Oct 12, 2021
Termination Date
Apr 11, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
72 units

Product description

Artis icono biplane with software VE20C-diagnostic imaging angiography system Model: 11327600

Reason for recall

Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap; 4. Unintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option

Distribution pattern

US Nationwide distribution.

Code information

Serial Numbers: 180062 180318 180054 180019 180065 180057 180041 180317 180034 180072 180349 180060 180337 180070 180048 180083 180369 180370 180345 180315 180343 180080 180301 180305 180067 180330 180040 180360 180333 180029 180352 180336 180316 180068 180363 180084 180334 180018 180350 180354 180046 180047 180335 180085 180035 180339 180357 180358 180361 180362 180081 180331 180021 180342 180321 180309 180356 180368 180341 180323 180329 180043 180049 180077 180086 180053 180055 180015 180014 180066 180082 180037 UDI: 04056869063317 Expanded Recall 12/20/21 S/N: 180056

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0516-2022
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Medical Solutions USA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Medical Solutions USA, Inc →

Source: openFDA record for Z-0516-2022. Report a data issue: hello@regdatalab.com.