Z-0523-2024 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-0523-2024 was initiated by Siemens Medical Solutions USA, Inc on Oct 17, 2023 and is designated Class II. Reason for recall: Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g… The recall status is ongoing. Affected quantity: 38 worldwide, 11 US.

Recall details

Product Type
Devices
Report Date
Dec 20, 2023
Initiation Date
Oct 17, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
38 worldwide, 11 US

Product description

ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100

Reason for recall

Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and switched on. During this time, system movement is only possible with the "override" function with very slow speed and without further collision protection. This may lead to temporary unavailability of the system for diagnostic and therapeutic usage.

Distribution pattern

US Nationwide - Worldwide Distribution: Domestic distribution to FL GA IA IL IN MI MN MO MT PA TX. Foreign distribution to Australia Canada Switzerland China Germany Denmark Egypt Spain Finland United Kingdom Israel Netherlands Sweden South Africa

Code information

UDI-DI 04056869295923 Serial Numbers: All SN's associated with material number 11328100

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0523-2024
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Medical Solutions USA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Medical Solutions USA, Inc →

Source: openFDA record for Z-0523-2024. Report a data issue: hello@regdatalab.com.