Z-0525-2013 Class I Terminated

Recalled by Praxair Inc. — Danbury, CT

FDA device recall Z-0525-2013 was initiated by Praxair Inc. on November 16, 2012 and is designated Class I. Reason for recall: Isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwise subject to significant physical impact. The recall status is terminated (terminated March 24, 2014). Affected quantity: 320,451 units (255,690 US ; 55,884 Canada; 8877 OUS).

Recall Details

Product Type
Devices
Report Date
January 2, 2013
Initiation Date
November 16, 2012
Termination Date
March 24, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
320,451 units (255,690 US ; 55,884 Canada; 8877 OUS)

Product Description

Praxair Vantage Grab n Go - Gas Cylinder pressure regulator Catalog Number: OX-M-AEGNGVNTG Usage: Regulator reduces pressure in oxygen-containing cylinder.

Reason for Recall

Isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwise subject to significant physical impact.

Distribution Pattern

Worldwide distribution: USA (nationwide) and countries of: Brazil, Canada, China, Columbia, Costa Rica, India, Paraguay and Uruguay.

Code Information

OX M-AEGNGVNTG (US and Canada) and WESPRX-9500 (Global) Cyclinders in service since Jan 2009. Units with a "T" decal (T-teflon) on the fill side of the valve have the rework completed.