Z-0525-2013 Class I Terminated

Recalled by Praxair Inc. — Danbury, CT

FDA device recall Z-0525-2013 was initiated by Praxair Inc. on Nov 16, 2012 and is designated Class I. Reason for recall: Isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwise subject to significant physical impact. The recall status is terminated (terminated Mar 24, 2014). Affected quantity: 320,451 units (255,690 US ; 55,884 Canada; 8877 OUS).

Recall details

Product Type
Devices
Report Date
Jan 2, 2013
Initiation Date
Nov 16, 2012
Termination Date
Mar 24, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
320,451 units (255,690 US ; 55,884 Canada; 8877 OUS)

Product description

Praxair Vantage Grab n Go - Gas Cylinder pressure regulator Catalog Number: OX-M-AEGNGVNTG Usage: Regulator reduces pressure in oxygen-containing cylinder.

Reason for recall

Isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwise subject to significant physical impact.

Distribution pattern

Worldwide distribution: USA (nationwide) and countries of: Brazil, Canada, China, Columbia, Costa Rica, India, Paraguay and Uruguay.

Code information

OX M-AEGNGVNTG (US and Canada) and WESPRX-9500 (Global) Cyclinders in service since Jan 2009. Units with a "T" decal (T-teflon) on the fill side of the valve have the rework completed.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0525-2013
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Praxair Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works