Z-0526-2018 Class II Terminated

Recalled by Angiodynamics, Inc. — Queensbury, NY

FDA device recall Z-0526-2018 was initiated by Angiodynamics, Inc. on Jun 5, 2017 and is designated Class II. Reason for recall: The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an… The recall status is terminated (terminated Sep 11, 2018). Affected quantity: 14,353 units in total.

Recall details

Product Type
Devices
Report Date
Feb 14, 2018
Initiation Date
Jun 5, 2017
Termination Date
Sep 11, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14,353 units in total

Product description

Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600, Model # (UPN) H7879006000, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

Reason for recall

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Distribution pattern

Worldwide distribution - US Nationwide

Code information

Lots: 15310411, 15320416, 15340428, 15350430, 15360434, 15380439, 15400449, 15420453, 15430454, 15440459, 15460466, 15470468, 15490476, 15500480, 15520487, 16010488, 16050494, 16070499, 16080502, 16100508, 16120512, 16140517, 16150520, 16170528, 16180530, 16190535, 16210540, 16230549, 16240550, 16250556, 16260561, 16270562, 16300568, 16310572, 16320573, 16350581, 16370590, 16380592, 16390595, 16400599, 16410603, 16420604, 16440611, 16450617, 17010645, 17020649, 17030653, 17040657, 17060662, 17070666, 17090673, 17100675, 17130686*, 17130687.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0526-2018
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Angiodynamics, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0526-2018. Report a data issue: hello@regdatalab.com.