Z-0527-2018 Class II Terminated

Recalled by Angiodynamics, Inc. — Queensbury, NY

Recall Details

Product Type
Devices
Report Date
February 14, 2018
Initiation Date
June 5, 2017
Termination Date
September 11, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14,353 units in total

Product Description

Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601, Model # (UPN) H7879006010, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

Reason for Recall

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Distribution Pattern

Worldwide distribution - US Nationwide

Code Information