Z-0527-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

Recall Details

Product Type
Devices
Report Date
November 26, 2025
Initiation Date
September 30, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
570 kits

Product Description

MEDLINE convenience kits labeled as: 1) GYN LAPAROTOMY HARPER PACK-LF, REF DYNJ0160684F; 2) MAJOR LITHOTOMY PACK-LF, REF DYNJ0425778O; 3) MINOR VAGINAL #76-RF, REF DYNJ27434R; 4) ABDOMINAL HYST PACK, REF DYNJ44848M; 5) DA VINCI HYSTERECTOMY, REF DYNJ44863Q; 6) VAGINAL HYSTERECTOMY PACK, REF DYNJ44900P; 7) URO GYN, REF DYNJ50047G; 8) ISS VAG HYST PACK, REF DYNJ50315F; 9) PK MAJOR GYN ST MICHAEL C, REF DYNJ60554B; 10) VAGINAL HYSTERECTOMY PACK, REF DYNJ67194B; 11) LAVH, REF DYNJ900298L; 12) LITHOTOMY-LF, REF DYNJ900721C; 13) GP-T A H PACK-LF, REF PHS396954006G; 14) GP-VAG HYSTERECTOMY PACK-LF, REF PHS396975006F.

Reason for Recall

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

Distribution Pattern

US Nationwide distribution.

Code Information

1) REF DYNJ0160684F: UDI/DI 10195327304287 (each) 40195327304288 (case), Lot Numbers: 23DDC245, 23GDC219; 2) REF DYNJ0425778O: UDI/DI 10198459082290 (each) 40198459082291 (case), Lot Numbers: 24LMF192; 3) REF DYNJ27434R: UDI/DI 10195327107727 (each) 40195327107728 (case), Lot Numbers: 24BBR389, 24CBH709; 4) REF DYNJ44848M: UDI/DI 10195327415952 (each) 40195327415953 (case), Lot Numbers: 23EDB911, 23FDC315, 23HDA180; 5) REF DYNJ44863Q: UDI/DI 10195327298227 (each) 40195327298228 (case), Lot Numbers: 23FDB122, 23GDC100; 6) REF DYNJ44900P: UDI/DI 10195327298548 (each) 40195327298549 (case), Lot Numbers: 23DDB824, 23EDA215, 23FDB287; 7) REF DYNJ50047G: UDI/DI 10195327107734 (each) 40195327107735 (case), Lot Numbers: 24BBE163, 24DBO682; 8) REF DYNJ50315F: UDI/DI 10195327535353 (each) 40195327535354 (case), Lot Numbers: 23KMF190; 9) REF DYNJ60554B: UDI/DI 10195327361747 (each) 40195327361748 (case), Lot Numbers: 23EBD224, 23KBE886; 10) REF DYNJ67194B: UDI/DI 10195327299699 (each) 40195327299690 (case), Lot Numbers: 23HBN435; 11) REF DYNJ900298L: UDI/DI 10195327554170 (each) 40195327554171 (case), Lot Numbers: 23LBF269, 23LBT933, 24BBT248; 12) REF DYNJ900721C: UDI/DI 10193489951608 (each) 40193489951609 (case), Lot Numbers: 23HBW996; 13) REF PHS396954006G: UDI/DI 10195327454159 (each) 40195327454150 (case), Lot Numbers: 23HDB088; 14) REF PHS396975006F: UDI/DI 10195327454166 (each) 40195327454167 (case), Lot Numbers: 23JDB593.