Z-0533-2026 Class II Ongoing
Recalled by Merit Medical Systems, Inc.
— South Jordan, UT
FDA device recall Z-0533-2026 was initiated by Merit Medical Systems, Inc. on Aug 4, 2025 and is designated Class II. Reason for recall: Hemostasis valve has a manufacturing defect and when used the valve cap may remain depressed, leading to potential leakage, which may result in procedure inconvenience and/or hemorrhage. The recall status is ongoing. Affected quantity: 217,141.
Recall details
- Product Type
- Devices
- Report Date
- Nov 26, 2025
- Initiation Date
- Aug 4, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 217,141
Product description
PhD Hemostasis Valve, REF: IN4802/K, IN4802/T, IN4918/A, IN7802/C, IN8802/C, IN9802, IN9802/A, K05-02166B, K05T-02138A, K12T-10798A, MAP800/B, MAP800/T, MAP801/B, MAP802/B, MAP802/CNB, MAP802/T, MAP804/B, MAP852/B, SKY1802.
Reason for recall
Hemostasis valve has a manufacturing defect and when used the valve cap may remain depressed, leading to potential leakage, which may result in procedure inconvenience and/or hemorrhage.
Distribution pattern
US: PA, AZ, LA, CO, FL, NC, SC, OH, AK, IL, IA, MN, TX, WI, AR, NY, CA, KS, NJ, MI, VA, KY, AL, MT, NE, MA, MO, TN, CT, MD, WA, DC, ME, OR, NH, IN, GA, UT, OK, HI, RI, ID, NM, MS, WV, SD, DE, NV, ND. OUS: United Arab Emirates, Australia, Italy, Spain, Sweden, Norway, Germany, France, Belgium, Netherlands, Thailand, New Zealand, Mexico, Puerto Rico, Canada, Korea, Republic of, South Africa, Qatar, Iceland, United Kingdom, Viet Nam, Slovakia, Guam, Oman, Portugal, Madagascar, Saudi Arabia, Finland, Cyprus, Switzerland, Taiwan, China, Japan
Code information
REF/UDI-DI/Lot(Expiration): IN4802/K/00884450638385/I3138883(20-FEB-2028), I3150031(27-FEB-2028), I3161209(10-MAR-2028), I3161210(08-APR-2028), I3161211(17-MAR-2028), I3172451(30-NOV-2026); IN4802/T/00884450860113/I3135188(30-NOV-2026, I3170171(10-MAR-2028; IN4918/A/00884450638521/I3146050(20-FEB-2028), I3200779(19-MAY-2028); IN7802/C/00884450498743/H3143753(30-NOV-2026), H3162387(14-MAR-2028), H3183068(31-MAR-2028), H3189446(29-FEB-2028), H3213617(31-MAY-2028); IN8802/C/00884450621417/H3141570(02-FEB-2028), H3143769(03-FEB-2028), H3143770(31-DEC-2026), H3143771(03-FEB-2028), H3143772(03-FEB-2028), H3149478(17-FEB-2028), H3149480(31-DEC-2026), H3149483(31-DEC-2026), H3155549(24-FEB-2028), H3162422(14-MAR-2028), H3162627(14-MAR-2028), H3168247(07-MAR-2028), H3183054(31-DEC-2026), H3200495(26-APR-2028), H3219087(10-JUN-2028), H3241812(31-MAY-2028); IN9802/00884450413463/H3141573(21-MAR-2028), H3183200(28-MAR-2028); IN9802/A/00884450021347/H3200469(26-APR-2028), H3210336(30-NOV-2026); K05-02166B/00884450499597/T3192808(24-FEB-2027), T3198985(11-AUG-2027), T3201276(27-FEB-2027), T3217490(02-SEP-2027); K05T-02138A/ 00884450095607 T3232335(15-AUG-2027); K12T-10798A/00884450544372/T3163727(21-MAY-2027), T3217266(20-FEB-2026); MAP800/B/ 00884450746714/I3124454(07-JAN-2028), I3174736(10-MAR-2028), I3186951(24-MAR-2028), I3191334(01-APR-2028), I3206933(21-APR-2028), I3209939(29-APR-2028), I3219784(12-MAY-2028), I3219792(12-MAY-2028), I3234557(19-MAY-2028), I3234558(26-MAY-2028); MAP800/T/00884450003985/I3162929(10-MAR-2028), I3162931(17-MAR-2028), I3186954(17-MAR-2028), I3186961(15-APR-2028); MAP801/B/00884450746721/I3146099 (24-MAR-2028), I3186946(21-APR-2028), I3209937(31-DEC-2026); MAP802/B/00884450746745/I3118322(21-JAN-2028), I311832407-JAN-2028), I3129725(06-FEB-2028), I3129726(06-FEB-2028), I3150110(20-FEB-2028), I3150111(20-FEB-2028), I3162940(27-FEB-2028), I3162942(03-MAR-2028), I3174729(03-MAR-2028), I3174730(17-MAR-2028), I3192937(15-APR-2028), I3200996(15-APR-2028), I3209935(29-APR-2028), I3215925(29-APR-2028), I3215926(30-APR-2028), I3215928(05-MAY-2028), I3219783(30-APR-2028), I3219785(19-MAY-2028), I3219786(19-MAY-2028), I3230013(26-MAY-2028); MAP802/CNB/00884450746752/I3174740(08-APR-2028); MAP802/T/00884450860939/I3162943(17-MAR-2028), I3192650(08-APR-2028), I3192654(22-APR-2028); MAP804/B/00884450746783/I3146101(31-OCT-2027), I3161426(31-DEC-2027); MAP852/B/00884450746820/I3187883(28-FEB-2028); SKY1802/00884450822500/H3079946(20-OCT-2027), H3141576(21-MAR-2028), H3141577(21-MAR-2028), H3149507(28-MAR-2028), H3149508(28-MAR-2028), H3155575(21-MAR-2028), H3155576(21-MAR-2028), H3155577(21-MAR-2028), H3155578(14-MAR-2028), H3168264(29-FEB-2028), H3168265(29-FEB-2028), H3182996(28-MAR-2028), H3183206(31-MAR-2028), H3183207(31-MAR-2027), H3189428(12-APR-2028), H3189451(12-APR-2028), H3200416(19-APR-2028), H3200475(19-APR-2028), H3219111(30-NOV-2026)
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0533-2026
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.