Z-0536-2014 Class II Terminated
FDA device recall Z-0536-2014 was initiated by Roche Diagnostics Operations, Inc. on Nov 25, 2013 and is designated Class II. Reason for recall: This issue is only relevant for customers importing xml files into the MagNA Pure 96 IVD instrument software for sample identification. There is a potential for sample mismatch when using the MagNA… The recall status is terminated (terminated Sep 4, 2014). Affected quantity: 27.
Recall Details
- Product Type
- Devices
- Report Date
- Dec 25, 2013
- Initiation Date
- Nov 25, 2013
- Termination Date
- Sep 4, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 27
Product Description
MagNA Pure 96 System Version 1.0 Software Version 2.0 Robotic workstation for automated isolation of nucleic acids from different kinds of crude sample material (whole blood, serum, blood cells, culture cells, tissue, bacteria, fungi) by the use of the specially designed MagNA Pure reagent kits, automated filling of different kinds of PCR reaction vessels.
Reason for Recall
This issue is only relevant for customers importing xml files into the MagNA Pure 96 IVD instrument software for sample identification. There is a potential for sample mismatch when using the MagNA Pure 96 IVD instrument (catalog number 06541089001). Importing an order file in xml format to the MagNA Pure 96 IVD instrument software may result in a wrong sequence of samples in the graphical overvi
Distribution Pattern
US Distribution including the states of AL, AZ, CA, CT, GA, NY, NC, NJ, OH, TX, VA and WA.
Code Information
Serial Numbers 2001, 2002, 2003, 2004, 2006, 2011, 2012, 2013, 2014, 2015, 2016, 2018, 2019, 2020, 2021, 2022, 2023, 2024, 2025, 2026, 2027, 2028, 4001, 4022, 4023, 4024, 4025,