Z-0540-2022 Class II Terminated

Recalled by DIXI MEDICAL USA — Plymouth, MI

FDA device recall Z-0540-2022 was initiated by DIXI MEDICAL USA on Dec 9, 2021 and is designated Class II. Reason for recall: Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag The recall status is terminated (terminated Sep 8, 2023). Affected quantity: 345 units.

Recall details

Product Type
Devices
Report Date
Feb 9, 2022
Initiation Date
Dec 9, 2021
Termination Date
Sep 8, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
345 units

Product description

The DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrode that is 0.8 mm in diameter with a rounded (hemispherical) tip. The electrode has multiple platinum/iridium contacts that are 2 mm long, with a total exploration length varying from 16 mm to 80.5 mm according to the electrode reference. Microdeep Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain.

Reason for recall

Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag

Distribution pattern

US Nationwide distribution.

Code information

All lots UDI:03664539000114

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0540-2022
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for DIXI MEDICAL USA? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for DIXI MEDICAL USA: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about DIXI MEDICAL USA →

Source: openFDA record for Z-0540-2022. Report a data issue: hello@regdatalab.com.