Z-0544-2021 Class II Terminated

Recalled by Philips North America Llc — Cambridge, MA

Recall Details

Product Type
Devices
Report Date
December 16, 2020
Initiation Date
November 18, 2020
Termination Date
July 22, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 total: 2 units US ; 14 OUS

Product Description

Philips IntelliVue X2 Multi measurement Server: M3002A (865039) - Product Usage: intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment and during patient transport inside and outside of the hospital environment.

Reason for Recall

Speakers on IntelliVue X2 and IntelliVue MP2 may fail, identified by the display of a Speaker Malfunct. INOP or by the absence of audible sound, may result in delayed treatment of the patient.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the state of SD and the countries of China, England, Germany, Hong Kong, Portugal.

Code Information

Serial Numbers: US: DE73607502 DE73607498 OUS: System Serial Number DE95038935 DE73608191 DE73608209 DE73601727 DE73604240 DE73601547 DE73601596 DE73604182 DE83617535 DE73608259 DE83617500 DE83617522 DE73601651