Z-0548-2024 Class I Ongoing

Recalled by Integra LifeSciences Corp. — Princeton, NJ

FDA device recall Z-0548-2024 was initiated by Integra LifeSciences Corp. on Oct 5, 2023 and is designated Class I. Reason for recall: Firm is expanding the existing recall RES 90457 for the Cerelink, initiated in June 2022, to include all CereLink ICP Extension Cables. The updated corrective action includes a design change to the… The recall status is ongoing. Affected quantity: 4941.

Recall details

Product Type
Devices
Report Date
Dec 20, 2023
Initiation Date
Oct 5, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4941

Product description

CereLink ICP Extension Cable (Extension Cable), Model No. 826845 Used with Codman CereLink Intracranial Pressure (ICP) Monitor, Model No. 826820/826820P. Cable is also provided with monitor as part of the same SKU.

Reason for recall

Firm is expanding the existing recall RES 90457 for the Cerelink, initiated in June 2022, to include all CereLink ICP Extension Cables. The updated corrective action includes a design change to the extension cable to address the "out of range readings" issue.

Distribution pattern

US distribution nationwide to the following states: AL AR AZ CA CO CT DC DE FL GA IA KY LA MA ME MN ND NE NH NJ NY OK PA RI TN TX UT VA WI WV OUS distribution to Australia/New Zealand, Canada, and Europe/Middle East/Africa.

Code information

UDI-DI 10381780520665 All lots distributed

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0548-2024
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0548-2024. Report a data issue: hello@regdatalab.com.