Z-0549-2021 Class II Terminated

Recalled by Canon Medical System, USA, INC. — Tustin, CA

FDA device recall Z-0549-2021 was initiated by Canon Medical System, USA, INC. on Nov 9, 2020 and is designated Class II. Reason for recall: A software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the… The recall status is terminated (terminated Dec 16, 2022). Affected quantity: 60 units.

Recall details

Product Type
Devices
Report Date
Dec 16, 2020
Initiation Date
Nov 9, 2020
Termination Date
Dec 16, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
60 units

Product description

Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L - Product Usage: intended to acquire and display cross-sectional volumes of the whole body, including the head.

Reason for recall

A software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing resulting in rescanning and reinjection of contrast medium.

Distribution pattern

US Nationwide distribution. OUS: Unknown

Code information

Serial Numbers: 8LA2052063, 8LA19Z2034 , 8LA19X2004, 8LA19Z2027, 8LA19Z2029, 8LA2052062, 8LA2042055, 8LA19Z2032, 8OA2052059, 8LA2042051, 8LA2032044, 8LA19X2002, 1LB1922175, 8LA2042050, 8LA19X2007, 8LA2012037, 8LA19Y2021, 8LA19Y2014, 8LA2042056, 8LA19Z2023, 8LA2052067, 8LA19Z2031, 8LA19Y2019, 8LA19Z2030, 8OA19Y2013, 8LA19Z2028, 8LA19Y2012, 8LA19Y2020, 8LA2032042, 8LA2052066, 8LA19X2003, 8LA19Y2022, 8LA19X2008, 8LA2042047, 8LA19Z2024, 8LA2032046, 8LA19Z2033, 8OA2042053, 8LA2012036, 8LA2042054, 8LA19Z2025, 8LA19Z2026, 8LA2022041, 8LA19X2010, 8LA2032043, 8OA19X2001, 8OA2052061, 8OA2042049, 8LA19Y2016, 8LA2022039, 8LA19Y2015, 8LA2012035, 8LA2032045, 8LA19Y2018, 8LA19X2005, 8LA19X2006, 8LA19X2009, 8LA19Y2011, 8LA19Y2017, 8LA2012038, 8LA2052060, 8LA2052058 , 8LA2052064, 8LA2052065, 8LA2052068, 8LA2022040, 8OA2042052, 8LA2042048, 8LA2052057

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0549-2021
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Canon Medical System, USA, INC.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Canon Medical System, USA, INC.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Canon Medical System, USA, INC. →

Source: openFDA record for Z-0549-2021. Report a data issue: hello@regdatalab.com.