Z-0554-2025 Class II Ongoing

Recalled by Shimadzu Corp. Analytical & Measuring Instruments Div. — Kyoto

Recall Details

Product Type
Devices
Report Date
December 4, 2024
Initiation Date
October 9, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
48

Product Description

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Reason for Recall

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.

Code Information

Model/UDI-DI/Serial Numbers: SIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16.