Z-0559-2024 Class II Ongoing

Recalled by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Best, N/A

FDA device recall Z-0559-2024 was initiated by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. on Nov 3, 2023 and is designated Class II. Reason for recall: Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur… The recall status is ongoing. Affected quantity: 3821 units.

Recall details

Product Type
Devices
Report Date
Dec 27, 2023
Initiation Date
Nov 3, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3821 units

Product description

Philips Azurion systems with monoplane fixed ceiling mounts: 722078 Azurion 7 M12 722079 Azurion 7 M20 722223 Azurion 7 M12 722224 Azurion 7 M20 722227 Azurion 5 M12 722228 Azurion 5 M20

Reason for recall

Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose

Distribution pattern

Worldwide - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, CANADA, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, IRAQ, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Latvia, Latvia, Republic of Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen.

Code information

Mdoel/Product Name/UDI-DI: 722078 Azurion 7 M12 (01)00884838085251(21); 722079 Azurion 7 M20 (01)00884838085268(21); 722223 Azurion 7 M12 (01)00884838099241(21); 722224 Azurion 7 M20 (01)00884838099258(21); 722227 Azurion 5 M12 (01)00884838099227(21); 722228 Azurion 5 M20 (01)00884838099234(21).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0559-2024
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. →

Source: openFDA record for Z-0559-2024. Report a data issue: hello@regdatalab.com.