Z-0562-2024 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 3, 2024
- Initiation Date
- October 16, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 172 kits
Product Description
(1) Nursing Skills kit, SKU #EDUC05038B, Case UPC 40195327019878, Non-sterile, For Training Purposes Only, Not For Human Use.
Reason for Recall
The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
Distribution Pattern
US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.
Code Information
(1) EDUC05038B - Lot numbers 22IDA181, 22FDB543, and 22ADB550, case UPC40195327019878;