Z-0566-2016 Class II Terminated
Recalled by Aesculap, Inc.
— Center Valley, PA
FDA device recall Z-0566-2016 was initiated by Aesculap, Inc. on Nov 17, 2015 and is designated Class II. Reason for recall: AAG received a customer complaint that corrosion had been found by a hospitals central sterile service department on the ball joint of an FC424R blade. Investigation revealed that the corrosion was… The recall status is terminated (terminated Oct 13, 2016). Affected quantity: 41.
Recall details
- Product Type
- Devices
- Report Date
- Jan 6, 2016
- Initiation Date
- Nov 17, 2015
- Termination Date
- Oct 13, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 41
Product description
Valve XS Atrium Retractor FC429R 52154756 The Valve XS Atrium Lift Retractor is used in heart surgery. The retractor is used for the temporary dilation of the left atrium in mitral valve surgery and for assistance in tricuspid valve surgery.
Reason for recall
AAG received a customer complaint that corrosion had been found by a hospitals central sterile service department on the ball joint of an FC424R blade. Investigation revealed that the corrosion was due to incorrect material used in the manufacturing of the disk spring of the ball joint by the supplier
Distribution pattern
US Distribution to state of: NY.
Code information
Product Information All lots affected Product Description FC424R Valve XS Atrium Lift Ret. Blade50X35 FC425R Valve XS Atrium Lift Ret. Blade60X35 FC426R Valve XS Atrium Lift Ret. Blade70X35 FC428R Valve XS Atrium Lift Ret. Blade50X45 FC429R Valve XS Atrium Lift Ret. Blade60X45 FC430R Valve XS Atrium Lift Ret. Blade70X45 FC431R Valve XS Atrium Lift Ret. Blade25X20 FC432R Valve XS Atrium Lift Ret. Blade25X25 FC433R Valve XS Atrium Lift Ret. Blade40X25 FC434R Valve XS Atrium Lift Ret. Blade50X25 FC435R Valve XS Atrium Lift Ret. Blade25X40 FC436R Valve XS Atrium Lift Ret. Blade25X50 FC420R Valve XS Atrium Lift Ret. Blade Complete Set
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0566-2016
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Aesculap, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works