Z-0567-2016 Class II Terminated
FDA device recall Z-0567-2016 was initiated by Becton Dickinson & Company on Dec 3, 2015 and is designated Class II. Reason for recall: Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a… The recall status is terminated (terminated Oct 19, 2018). Affected quantity: 29,815,600 units in total.
Recall details
- Product Type
- Devices
- Report Date
- Jan 6, 2016
- Initiation Date
- Dec 3, 2015
- Termination Date
- Oct 19, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 29,815,600 units in total
Product description
BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 2.0 mL (Pink) BD Hemogard Additive: K2EDTA (spray dried), 3.6mg Sterile IVD BD, Made in USA The Vacutainer Brand PLUS (Plastic) Tube with EDTA Vacutainer¿ Brand Serum Tube are evacuated blood collection tubes which provide a means of collecting, transporting, separating and processing blood in a plastic tube.
Reason for recall
Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
Distribution pattern
Worldwide Distribution: US (Nationwide) and the countries of: European Union, Peoples Republic of China, Japan and Republic of Korea.
Code information
Catalog Reference #367842; Lot numbers 5089731 and 5175923