Z-0571-2025 Class II Ongoing

Recalled by LivaNova Deutschland GmbH — Munich

FDA device recall Z-0571-2025 was initiated by LivaNova Deutschland GmbH on October 18, 2024 and is designated Class II. Reason for recall: Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may oc… The recall status is ongoing. Affected quantity: 38.

Recall Details

Product Type
Devices
Report Date
December 4, 2024
Initiation Date
October 18, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
38

Product Description

Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software

Reason for Recall

Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.

Distribution Pattern

US Nationwide distribution in the states of CA, VA, GA, IN, NJ, FL, CO, MS, NV.

Code Information

UDI-DI: 04033817903062. Software Version: 1.5. Serial Numbers: 49PK00436, 49PK00108, 49PK00374, 49PK00479, 49PK00480, 49PK00481, 49PK00093, 49PK00091, 49PK00092, 49PK00163, 49PK00497, 49PK00498, 49PK00219, 49PK00223, 49PK00224, 49PK00225, 49PK00226, 49PK00241, 49PK00255, 49PK00292, 49PK00291, 49PK00293, 49PK00490, 49PK00487, 49PK00488, 49PK00278, 49PK00274, 49PK00277, 49PK00275, 49PK00276, 49PK00273, 49PK00457, 49PK00458, 49PK00459, 49PK00455, 49PK00456, 49PK00290, 49PK00288