Z-0073-2024 Class II Ongoing
FDA device recall Z-0073-2024 was initiated by LivaNova Deutschland GmbH on Aug 3, 2023 and is designated Class II. Reason for recall: The firm's labeling (Quick Reference Guide) contains incorrect information in the maintenance schedule for the Heater-Cooler device. The recall status is ongoing. Affected quantity: 37 systems.
Recall details
- Product Type
- Devices
- Report Date
- Oct 18, 2023
- Initiation Date
- Aug 3, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 37 systems
Product description
3T Heater Cooler System, REF: 16-02-85
Reason for recall
The firm's labeling (Quick Reference Guide) contains incorrect information in the maintenance schedule for the Heater-Cooler device.
Distribution pattern
US Nationwide distribution in the states of CA, IN, GA, NC, TX, AZ, FL, MN, NJ, CO, SC.
Code information
UDI: (01)04033817900528/ Serial Numbers: 20T31334, 20T31336, 20T31366, 20T31391, 20T31375, 20T31322, 20T31348, 20T31345, 20T31362, 20T31363, 20T31364, 20T31407, 20T31531, 20T31532, 20T31534, 20T31367, 20T31368, 20T31369, 20T31370, 20T31371, 20T31372, 20T31373, 20T31386, 20T31387, 20T31388, 20T31389, 20T31390, 20T31550, 20T31551, 20T31552, 20T31553, 20T31543, 20T31544, 20T31536, 20T31346, 20T31347, 20T31361, 20T31365