Z-2439-2020 Class II Terminated

Recalled by LivaNova Deutschland GmbH — Munich, N/A

FDA device recall Z-2439-2020 was initiated by LivaNova Deutschland GmbH on Aug 26, 2019 and is designated Class II. Reason for recall: The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a… The recall status is terminated (terminated Apr 22, 2021). Affected quantity: 8.

Recall details

Product Type
Devices
Report Date
Jul 8, 2020
Initiation Date
Aug 26, 2019
Termination Date
Apr 22, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8

Product description

Control Panel REF - 60-02-50: LivaNova SCP Pump Control Panel S5/C5. REF 60-02-50, IPX2, RxOnly, UDI: 04033817900993

Reason for recall

The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder. The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.

Distribution pattern

US: FL, GA, IN, KY, MI, MO, NY, OH, PA, SC, TN, TX, VA OUS: Canada - Uzbekistan - Turkey - Tunisia - Taiwan - South Africa - Serbia - Saudi Arabia - Russia - Poland - Mexico - Korea(Rep. of) - India - Hong Kong - Colombia - China - Brazil - Bahrain - Argentina - Algeria - Italy - Germany

Code information

Serial Number: 60S14735 60S14877 60S14715 60S14716 60S14736 60S14755 60S14707 60S14708

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2439-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for LivaNova Deutschland GmbH? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for LivaNova Deutschland GmbH: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about LivaNova Deutschland GmbH →

Source: openFDA record for Z-2439-2020. Report a data issue: hello@regdatalab.com.