Z-2440-2020 Class II Terminated

Recalled by LivaNova Deutschland GmbH — Munich, N/A

Recall Details

Product Type
Devices
Report Date
July 8, 2020
Initiation Date
August 26, 2019
Termination Date
April 22, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
24

Product Description

Control Panel REF - 60-02-15: LivaNova SCP Pump Control Panel. REF 60-02-15, IPX2, RxOnly, UDI: 04033817900986

Reason for Recall

The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder. The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.

Distribution Pattern

US: FL, GA, IN, KY, MI, MO, NY, OH, PA, SC, TN, TX, VA OUS: Canada - Uzbekistan - Turkey - Tunisia - Taiwan - South Africa - Serbia - Saudi Arabia - Russia - Poland - Mexico - Korea(Rep. of) - India - Hong Kong - Colombia - China - Brazil - Bahrain - Argentina - Algeria - Italy - Germany

Code Information