Z-0575-2025 Class II Ongoing

Recalled by Orthoscan, Inc. — Scottsdale, AZ

FDA device recall Z-0575-2025 was initiated by Orthoscan, Inc. on Aug 8, 2024 and is designated Class II. Reason for recall: Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with… The recall status is ongoing. Affected quantity: 3.

Recall details

Product Type
Devices
Report Date
Dec 11, 2024
Initiation Date
Aug 8, 2024
Voluntary/Mandated
FDA Mandated
Product Quantity
3

Product description

Ziehm Vision system, x-ray, tomography, computed.

Reason for recall

Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.

Distribution pattern

U.S. (including Puerto Rico).

Code information

UDI: EZIEZIEHMVISIONFD1 Serial Numbers: 9246, 90155, 90416.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0575-2025
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Orthoscan, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0575-2025. Report a data issue: hello@regdatalab.com.