Z-0578-2018 Class II Terminated

Recalled by LITTLE RAPIDS CORPORATION — Green Bay, WI

FDA device recall Z-0578-2018 was initiated by LITTLE RAPIDS CORPORATION on Dec 27, 2017 and is designated Class II. Reason for recall: The secondary supplier provided material was within specification but did not meet our design needs compared to the primary manufacturer. The product is designed for patient transfer, not lifting… The recall status is terminated (terminated Jul 30, 2018). Affected quantity: 153 cases, 30 units in each case.

Recall details

Product Type
Devices
Report Date
Feb 21, 2018
Initiation Date
Dec 27, 2017
Termination Date
Jul 30, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
153 cases, 30 units in each case

Product description

EMS products from Graham Medical, SnugFit(R) Expandable Mattress Cover, Code 72930, UPC 1 07 47036 72930 3

Reason for recall

The secondary supplier provided material was within specification but did not meet our design needs compared to the primary manufacturer. The product is designed for patient transfer, not lifting. However, we recognize that the product may be utilized outside of its intended design scope, and lifting may occur. Lifting could lead tearing of the equipment cover and the potential for patient falling through.

Distribution pattern

TX, CT, TN, MO, WI, FL, SD, MS, CA, MS

Code information

Lot codes: 17871, 17971, 18071, and 18971

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0578-2018
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for LITTLE RAPIDS CORPORATION? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0578-2018. Report a data issue: hello@regdatalab.com.