Z-0579-2026 Class II Ongoing

Recalled by Olympus Corporation of the Americas — Center Valley, PA

Recall Details

Product Type
Devices
Report Date
December 3, 2025
Initiation Date
October 14, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7023 units

Product Description

Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

Reason for Recall

Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.

Distribution Pattern

US Nationwide distribution.

Code Information

Model/Catalog Number: TJF-Q190V; UDI: 04953170405563 / 04953170452024; Serial Numbers: All;