Z-1639-2026 Class II Ongoing

Recalled by Olympus Corporation of the Americas — Center Valley, PA

Recall Details

Product Type
Devices
Report Date
April 1, 2026
Initiation Date
February 25, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
164 units

Product Description

Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.

Reason for Recall

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Distribution Pattern

Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.

Code Information

Model: 3006; UDI: 00821925035898; Lot Numbers: All unexpired;