Z-1638-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 1, 2026
- Initiation Date
- February 25, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 106 units
Product Description
Brand Name: Olympus Everest Bipolar Cutting Forceps, 5 mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3005 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.
Reason for Recall
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Distribution Pattern
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
Code Information
Model: 3005; UDI: 00821925035881; Lot Number: All unexpired