Z-0582-2016 Class II Terminated

Recalled by Vidco, Inc. — Beaverton, OR

FDA device recall Z-0582-2016 was initiated by Vidco, Inc. on Nov 8, 2015 and is designated Class II. Reason for recall: Testing at customer site showed unit Remote Patient Monitoring System MDP2040-0100 in a continuous trap condition, not allowing system to reset and reboot. Two customers recently complained of the… The recall status is terminated (terminated Aug 16, 2016). Affected quantity: 59 units.

Recall details

Product Type
Devices
Report Date
Jan 13, 2016
Initiation Date
Nov 8, 2015
Termination Date
Aug 16, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
59 units

Product description

Remote Patient Monitoring System, MDP 2000 Series, Part Numbers: MDP2040-0100, PGM340R3, or PGM340R4. The MDP2000 is intended for use in conjunction with patient monitoring equipment. The MDP2000 is not patient connected Examples of areas where the MDP200 is used include ICU, CCU, PACU, Emergency, Telemetry Step-down and other areas where patient's physiological information are to be observed at one or more locations. Typical information displayed includes ECG and blood pressure waveforms and numeric values which as heart rate and systolic, mean and diastolic pressure. This device acts as the Central Monitor Station, it permits from one (1) to sixteen (16) bedside monitor(s) to be connected simultaneously.

Reason for recall

Testing at customer site showed unit Remote Patient Monitoring System MDP2040-0100 in a continuous trap condition, not allowing system to reset and reboot. Two customers recently complained of the system freezing and it could only be restarted if the user re-applied power.

Distribution pattern

US Nationwide distribution in the states of AZ, CA, MD, NM, NJ, and OH.

Code information

Serial numbers: 6533-6537, 6561, 6566-6581, 6539, 6540, 6558-6560, 6562, 6508-6528, 6563-6565, 6529, 6531, 6532, 6541, 6551, 6556, and 6557.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0582-2016
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Vidco, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0582-2016. Report a data issue: hello@regdatalab.com.