Z-0584-2016 Class II Terminated

Recalled by Smith & Nephew, Inc. — Memphis, TN

FDA device recall Z-0584-2016 was initiated by Smith & Nephew, Inc. on Nov 16, 2015 and is designated Class II. Reason for recall: The 16 devices distributed in the United States were incorrectly labeled as the monoblock Birmingham Hip Modular Head, with labeling not cleared for marketing in the US. The actual devices were… The recall status is terminated (terminated Aug 25, 2016). Affected quantity: 16 units in US.

Recall details

Product Type
Devices
Report Date
Jan 13, 2016
Initiation Date
Nov 16, 2015
Termination Date
Aug 25, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 units in US

Product description

BIRMINGHAM HIP RESURFACING MODULAR HEAD, REF numbers: 74121238, MODULAR HEAD 38 MM -8MM 74121242, MODULAR HEAD 42 MM -8MM 74121246, MODULAR HEAD 46 MM -8MM 74121250, MODULAR HEAD 50 MM -8MM 74121254, MODULAR HEAD 54 MM -8MM 74121258, MODULAR HEAD 58 MM -8MM 74121338, MODULAR HEAD 3 8MM -4 (SHORT) 74121342, MODULAR HEAD 42 MM -4 (SHORT) 74121346, MODULAR HEAD 46 MM -4 (SHORT) 74121350, MODULAR HEAD 50 MM -4 (SHORT) 74121354, MODULAR HEAD 54 MM -4 (SHORT) 74121358, MODULAR HEAD 58 MM -4 (SHORT) 74121438, MODULAR HEAD 38 MM +0 (MED) 74121442, MODULAR HEAD 42 MM +0 (MED) 74121446, MODULAR HEAD 46 MM +0 (MED) 74121450, MODULAR HEAD 50 MM +0 (MED) 74121454, MODULAR HEAD 54 MM +0 (MED) 74121458, MODULAR HEAD 58 MM +0 (MED) 74121538, MODULAR HEAD 38 MM +4 (LONG) 74121542, MODULAR HEAD 42 MM +4 (LONG) 74121546, MODULAR HEAD 46 MM +4 (LONG) 74121550, MODULAR HEAD 50 MM +4 (LONG) 74121554, MODULAR HEAD 54 MM +4 (LONG) 74121558, MODULAR HEAD 58 MM +4 (LONG)

Reason for recall

The 16 devices distributed in the United States were incorrectly labeled as the monoblock Birmingham Hip Modular Head, with labeling not cleared for marketing in the US. The actual devices were monoblock Modular Femoral Head for hemiarthroplasty use.

Distribution pattern

Worldwide distribution. US nationwide, France, UAE, Austria, Australia, Belgium, CANADA, Switzerland, Germany, Denmark, Spain, Finland, UK, Hong Kong, ISRAEL, India, Italy, Japan, South Korea, Netherlands, Norway, Poland, PORTUGAL, Sweden, SINGAPORE, Thailand, Turkey, and SOUTH AFRICA.

Code information

All codes

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0584-2016
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Smith & Nephew, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Smith & Nephew, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Smith & Nephew, Inc. →

Source: openFDA record for Z-0584-2016. Report a data issue: hello@regdatalab.com.