Z-0588-2026 Class II Ongoing

Recalled by Pivotal Health Solutions, Inc. — Watertown, SD

FDA device recall Z-0588-2026 was initiated by Pivotal Health Solutions, Inc. on Oct 9, 2025 and is designated Class II. Reason for recall: The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and… The recall status is ongoing. Affected quantity: 464 units.

Recall details

Product Type
Devices
Report Date
Dec 3, 2025
Initiation Date
Oct 9, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
464 units

Product description

Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table

Reason for recall

The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.

Distribution pattern

Distribution to US nationwide, Canada and Japan.

Code information

All serial numbers.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0588-2026
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Pivotal Health Solutions, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Pivotal Health Solutions, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Pivotal Health Solutions, Inc. →

Source: openFDA record for Z-0588-2026. Report a data issue: hello@regdatalab.com.