Z-0589-2016 Class II Terminated
FDA device recall Z-0589-2016 was initiated by Maquet Cardiovascular Us Sales, Llc on Nov 13, 2015 and is designated Class II. Reason for recall: Failure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board… The recall status is terminated (terminated Apr 7, 2017). Affected quantity: 100 Units (distributed in United States).
Recall details
- Product Type
- Devices
- Report Date
- Jan 13, 2016
- Initiation Date
- Nov 13, 2015
- Termination Date
- Apr 7, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 100 Units (distributed in United States)
Product description
Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W Product Usage: The Heater Cooler Unit (HCU) 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.
Reason for recall
Failure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board. Board failure typically occurs when the compressor is switched on (prior to patient use); however, it can occur during patient use. This can shut down the compressor or the entire HCU 30 Type 2 unit.
Distribution pattern
United States Nationwide Distribution
Code information
HCU 30 Base Unit 100 - 120 V, Model Number 704629/70103.4371 HCU 30 Base Unit 200 - 240 V, Model Number 704631/70103.4642 HCU 30 Base Unit Canadian/US, Model Number 705501/70103.4653 Heater Cooler Unit 30 with Serial Number <93101480 The affected power supply boards have serial numbers less than 107006001.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0589-2016
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026