Z-0589-2016 Class II Terminated

Recalled by Maquet Cardiovascular Us Sales, Llc — Wayne, NJ

FDA device recall Z-0589-2016 was initiated by Maquet Cardiovascular Us Sales, Llc on Nov 13, 2015 and is designated Class II. Reason for recall: Failure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board… The recall status is terminated (terminated Apr 7, 2017). Affected quantity: 100 Units (distributed in United States).

Recall details

Product Type
Devices
Report Date
Jan 13, 2016
Initiation Date
Nov 13, 2015
Termination Date
Apr 7, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
100 Units (distributed in United States)

Product description

Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W Product Usage: The Heater Cooler Unit (HCU) 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

Reason for recall

Failure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board. Board failure typically occurs when the compressor is switched on (prior to patient use); however, it can occur during patient use. This can shut down the compressor or the entire HCU 30 Type 2 unit.

Distribution pattern

United States Nationwide Distribution

Code information

HCU 30 Base Unit 100 - 120 V, Model Number 704629/70103.4371 HCU 30 Base Unit 200 - 240 V, Model Number 704631/70103.4642 HCU 30 Base Unit Canadian/US, Model Number 705501/70103.4653 Heater Cooler Unit 30 with Serial Number <93101480 The affected power supply boards have serial numbers less than 107006001.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0589-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0589-2016. Report a data issue: hello@regdatalab.com.